توضیحات

TU5 Series® TU5200 Laboratory Laser Turbidimeter without RFID, EPA Version
Product Number: LPV442.99.01012
کدورت سنج
Specifications
Technical Attributes
Accuracy ±2 % plus 0.01 NTU from 0 – 40 NTU;
±10 % of reading from 40 – 1000 NTU based on Formazin primary standard (at 25 °C)
Calibration Method For Formazin and Stablcal:
20 NTU from 0 to 40 NTU; at 20 FNU and 600 NTU for full range

Custom calibration for up to 6-point calibrations

Degrees 0 – 100 mg/L
SDVB 0 – 1000 NTU
Certifications CE compliant
US FDA accession number: 1420493-000 EPA version, 1420492-000 ISO version
Complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)
Australian ACMA Marking
Communication Link2sc
Dimensions (H x W x D) 195 mm x 409 mm x 278 mm
Display 7″ WVGA (800 x 480) color TFT-LCD with touch screen
Enclosure Construction PC/ABS Blend Plastic
Light Source Class 2 laser product, with embedded 650 nm (EPA 0.43 mW) or Class 1 laser product, with embedded 850 nm (ISO), max. 0.55 mW (complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)
Model TU5200 EPA
Operating Humidity 80% at 30 °C (non condensing)
Operating Temperature Range 10 – 40 °C (50 – 104 °F)
Power Requirements (Hz) 50/60 Hz
Power Requirements (Voltage) 100 – 240 VAC
Range EPA:
0 – 700 NTU / FNU / TE/F / FTU
0 – 100 mg/L
0 – 175 EBC

ISO:
0 – 1000 NTU / FNU / TE/F / FTU
0 – 100 mg/L
0 – 250 EBC
Regulatory EPA
Repeatability 40 NTU: Better than 3.5% of reading on Formazin at 25 °C (77 °F)
Resolution 0.0001 NTU / FNU / TE/F / FTU / EBC / mg/L
Response Time Signal averaging off: 7 seconds
Signal averaging on: 10 seconds (when averaging time is 5 seconds)
Sample Temperature 4 – 70 °C (39 – 158 °F)
Source Lamp EPA Version – 650 nm Class 2 Laser
Storage Conditions -30 – 60 °C (-22 – 140 °F)
Stray Light <10 mNTU
Units NTU, FNU, TE/F, FTU, EBC;

mg/L if calibrated with Degrees calibration curve
Warranty 12 months
Weight 2.4 kg (5.29 lbs.)
What's included? TU5200 benchtop turbidimeter, user manual, Stablcal kit (10, 20 and 600 NTU), 2 sample vials, power supply, dust cover, vial stand

دیدگاهها

هیچ دیدگاهی برای این محصول نوشته نشده است.

اولین نفری باشید که دیدگاهی را ارسال می کنید برای “کدورت سنج TU5 Series® TU5200 Laboratory Laser Turbidimeter without RFID, EPA Version”